DISCLAIMER: The information provided herein is solely for informational purposes. Customers must be aware of the suitability of this product for their application, and consider that variable factors such as Manufacturer, Product Category, Date Codes, Pictures and Descriptions may differ from available inventory.

42510A V3.2 9/08
7
Emissivity and IR Measurement Theory
IR Thermometers measure the surface temperature of an object. The thermometer’s optics
sense emitted, reflected, and transmitted energy. The thermometer’s electronics translate
the information into a temperature reading which is then displayed on the LCD.
The amount of IR energy emitted by an object is proportional to an object's temperature
and its ability to emit energy. This ability is known as emissivity and is based upon the
material of the object and its surface finish. Emissivity values range from 0.1 for a very
reflective object to 1.00 for a flat black finish. For the Model 42510A, the emissivity is
adjustable from 0.1 to 1.00. Most organic materials and painted or oxidized surfaces have
an emissivity factor of 0.95. When in doubt, set the emissivity to 0.95.
Emissivity Factors for Common Materials
Material under test Emissivity Material under test Emissivity
Asphalt 0.90 to 0.98 Cloth (black) 0.98
Concrete 0.94 Skin (human) 0.98
Cement 0.96 Leather 0.75 to 0.80
Sand 0.90 Charcoal (powder) 0.96
Soil 0.92 to 0.96 Lacquer 0.80 to 0.95
Water 0.92 to 0.96 Lacquer (matt) 0.97
Ice 0.96 to 0.98 Rubber (black) 0.94
Snow 0.83 Plastic 0.85 to 0.95
Glass 0.90 to 0.95 Timber 0.90
Ceramic 0.90 to 0.94 Paper 0.70 to 0.94
Marble 0.94 Chromium Oxides 0.81
Plaster 0.80 to 0.90 Copper Oxides 0.78
Mortar 0.89 to 0.91 Iron Oxides 0.78 to 0.82
Brick 0.93 to 0.96 Textiles 0.90
Battery Replacement
When the low battery symbol appears on the display, replace the
meter’s battery (9V). The battery compartment is located behind the
panel that surrounds the meter’s trigger. Open the compartment by
pulling the panel down from the trigger area. Replace the 9V battery
and close the battery compartment cover.
42510A V3.2 9/08
8
Specifications
Range / Resolution 0.1° resolution from -58.0 to 999.9°F (-50.0 to 650.0°C);
1° resolution from 1000°F to 1200°F
Accuracy -58°F to -31°F (-50°C to -35°C) ±9°F/4°C
-31°F to 28°F (-35°C to -2°C) ±5°F/3°C
28°F to 200°F (-2°C to 94°C) ±3°F/2.5°C
200°F to 400°F (94°C to 204°C) ±(1.0%rdg + 2°F/1°C)
400°F to 800°F (204°C to 426°C) ±(1.5%rdg + 2°F/1°C)
800°F to 1200°F (426°C to 650°C) ±(2.0%rdg +2°F/1°C)
Note: Accuracy is specified for the following ambient temperature
range: 64 to 82°F (18 to 28°C)
Emissivity 0.10 to 1.00 adjustable
Field of View D/S = Approx. 12:1 ratio (D = distance; S = spot or target)
Laser pointer Class 2 laser < 1mW power; Wavelength: 630 to 670nm
IR Spectral response
8 to 14 μm
Repeatability ± 0.5% of reading or ± 1.8
o
F/1
o
C; whichever is greater
General Specifications
Display Backlit LCD display with function indicators
Response time 300ms
Over range indication “---------“
Operating Temperature 32
o
F to 122
o
F (0
o
C to 50
o
C)
Operating Humidity 10% to 90%RH operating, <80%RH storage.
Storage Temperature 14 to 140
o
F (-10 to 60
o
C)
Power Supply 9V battery
Automatic Power Off 7 seconds, with LOCK to disable
Weight 6.4 oz. / 180g
Dimensions 3.2 x 1.6 x 6.3” (82 x 42 x 160mm)
42510A V3.2 9/08
9
Support line (781) 890-7440
Technical support: Extension 200; E-mail: support@extech.com
Repair & Returns: Extension 210; E-mail: repair@extech.com
Product specifications subject to change without notice
For the latest version of this User’s Guide, Software updates, and other
up-to-the-minute product information, visit our website: www.extech.com
Extech Instruments Corporation, 285 Bear Hill Rd., Waltham, MA 02451
Calibration and Repair Services
Extech offers repair and calibration services for the products we sell. Extech also
provides NIST certification for most products. Call the Customer Service Department for
information on calibration services available for this product. Extech recommends that
annual calibrations be performed to verify meter performance and accuracy.